Code No. 13: Power Supply Interruption

Problem Description:

A momentary power failure or short circuit occurred during device operation — most often at the beginning of the sterilization cycle.

The issue may be caused by:

  • A short circuit in the autoclave or power outlet
  • Faulty electrical wiring in the installation site
  • Internal faults in components such as the valve wiring harness or heating elements

 

How to solve it:
  • Restart the device:
    • Tap a few times on the display until the Restart button appears.
    • Press the Restart button.
    • If a code is required, enter 0000.
    • If the Restart button does not appear, switch off the device manually, wait at least 20 seconds, and then switch it on again.
  • Ensure the autoclave is plugged directly into a working outlet
  • Try connecting it to a different socket to rule out an issue with the current one
  • Consult a qualified electrician to verify the electrical installation on-site.


If the issue persists, please contact Enbio Service by clicking the button below to fill out the service form.