Equalizing pressure

Problem description:

Pressure stabilization when the device is idle — equalization to atmospheric pressure.

Possible Causes:

  • A sterilization pouch may have fallen out of the tray and become lodged at the back of the chamber, blocking the drain outlet
  • A malfunctioning control board, which misreads pressure levels and falsely triggers this message

 

How to solve it:

This message may appear occasionally due to normal pressure equalization after a cycle. However, if it occurs frequently:

  • Inspect the chamber to ensure no pouch has slipped behind the tray and blocked the outlet and remove it if found.
  • Restart the device and observe behavior.

If the message persists or occurs without clear reason, please contact Enbio Service by clicking the button below to fill out the service form.