Code No. 10: Chamber Heating Issue

Problem Description:

The autoclave was unable to reach the required temperature on the chamber walls.

Several components may be responsible for this issue, including:

  • Defective heating elements
  • A faulty power relay on the control board
  • Issues with the power supply module
  • Damaged wiring harnesses
  • A malfunctioning mainboard
  • Improper electrical connection or insufficient power supply One frequent cause is connecting the autoclave to a power strip (multi-socket) that also supplies other devices. This can lead to voltage drops and unstable operation.

How to solve it:
  • Ensure the autoclave is connected directly to a dedicated power outlet, not shared with other devices.
  • Avoid using multi-socket extension strips.
  • Check that the mains voltage is stable and within specification (consult a certified electrician if needed).

If the problem persists, please contact Enbio Service by clicking the button below to fill out the service form