Code No. 5: Sterilization Pressure Too Low

Problem Description:

The pressure during the sterilization process is too low.

The most common cause is that the blue water supply hose is not fully submerged in water.

This can happen if:

  • The hose is incorrectly positioned inside the water container
  • The container is empty
  • The hose is damaged


If this is not the issue, the problem may be due to :

  • A clogged or worn-out Magic Filter
  • A worn-out water pump
  • An internal leak in the autoclave.
How to solve it:
  • Make sure the blue hose is fully submerged in water inside the supply container (remember to use the proper water).
  • Add a sinker to keep the hose submerged.
  • Inspect the hose for damage and replace if necessary.
  • Replace the water container if it is empty.
  • Check the Magic Filter and replace it if it appears used.
  • After making the adjustments, restart the autoclave.


If the problem persists, please contact Enbio Service by clicking the button below to fill out the service form.